missing translation for 'onlineSavingsMsg'
Learn More

Thermo Scientific™ Barnstead™ GenPure™ xCAD Plus Ultrapure Water Purification System, with Installation

Simultaneously dispense Type 1 ultrapure water from up to three dispensers using the Thermo Scientific™ Barnstead™ GenPure™ xCAD Plus system.

11440.00 € - 13000.00 €

Specifications

Conductivity 0.055µS/cm
Bacterial Content <0.1 CFU/mL
Feedwater Connector 3/4 in.
Feed Water Monitoring Integrated
Flow Rate Up to 2 L/min.
View More Specs

Products 6
Product Code Brand Model Price Quantity & Availability  
Product Code Brand Model Price Quantity & Availability  
16450841
View Documents
Thermo Scientific™
50136152INS
Barnstead GenPure xCAD Plus UV (bench version)
11440.00 €
Each
Estimated Shipment: 28-02-2025
to see stock.
Add to basket
 
16440841
View Documents Promotion Details
Thermo Scientific™
50136151INS
GenPure xCAD Plus UV/UF (bench version)
12168.00 €
Each
Estimated Shipment: 28-02-2025
to see stock.
Add to basket
 
16460841
View Documents
Thermo Scientific™
50136153INS
Barnstead GenPure xCAD Plus UV - TOC (bench version)
12272.00 €
Each
Estimated Shipment: 28-02-2025
to see stock.
Add to basket
 
16430841
View Documents
Thermo Scientific™
50136146INS
Barnstead GenPure xCAD Plus UV/UF - TOC (bench version)
13000.00 €
Each
Estimated Shipment: 28-02-2025
to see stock.
Add to basket
 
16470841
View Documents
Thermo Scientific™
50136169INS
Barnstead GenPure xCAD Plus UV/UF (wall version)
12168.00 €
Each
Estimated Shipment: 28-02-2025
to see stock.
Add to basket
 
16400851
View Documents Promotion Details
Thermo Scientific™
50136172INS
Barnstead GenPure xCAD Plus UV/UF - TOC (wall version)
13000.00 €
Each
Estimated Shipment: 28-02-2025
to see stock.
Add to basket
 
Description

Description

Simultaneously dispense Type 1 ultrapure water from up to three dispensers using the Thermo Scientific™ Barnstead™ GenPure™ xCAD Plus system.

All GenPure systems exceed international standards ASTM Type I, ISO 3696 Grade 1, ASTM D1193 and CLSI-CLRW.

xCAD (Extended Control and Dispenser):

  • System provides more flexibility for ultrapure water dispensing with an action radius of 32 in. (80cm).
  • Ergonomic dispensing arm provides precise volume dosing at the press of a button.
  • Flexible dispenser offers positions including turned, height adjusted or extended toward your vessel.
  • One-of-a-kind design provides access to all system controls at your fingertips.

Simultaneously dispense from up to three dispensers:

  • Each system ships with a single xCAD Plus remote dispenser.
  • For increased flexibility, add one or two more xCAD Plus dispensers.

Integrated feed water monitoring:

  • Additional measuring cell monitors feed water conductivity.
  • If the quality of the feed water drops below the set value, a fault message will immediately be displayed.

Online TOC measurement with UV intensity monitoring (available on systems with optional TOC only):

  • Real-time TOC monitoring offers continual measurement of organic substances present in the water.
  • Continuous monitoring of UV intensity removes the risk of false TOC values due to diminished UV intensity.
  • Fault message alerts user when the UV lamp intensity is no longer sufficient for accurate TOC measurements.

UV Photo-oxidation 185/254 nm:

  • High performance UV assembly reliably reduces the content of microorganisms and their metabolites.
  • UV oxidation also reduces organic compounds in the water to ultra-low levels.

Leak detection:

  • An error message is displayed when internal leaks are detected.

Ready to use:

  • Feed water pressure switch, wall bracket, UV lamp, ultrapure cartridge, sterile filter, all are included in one part number.

Measurement of conductivity/resistivity:

  • Conductivity cells are carefully calibrated prior to each measurement via built-in reference resistance with cell constants at 0.01cm-1.
  • Temperature measurements are made by a platinum chip sensor with ± 0.1°C accuracy.

Quick-connect cartridge replacement:

  • Aquastop quick-connect capability enables cartridge replacement within seconds even mid-operation.

Exact dosing:

  • Electronic dosing yields fully automatic volume control of 0.01-65 L with <2% accuracy.

GLP-compliant documentation:

  • Real-time clock and code-protected operating system prevents unauthorized changes to system settings.
  • RS-232 interface with adjustable send-interval offers safe data transfer of all measured data, faults, date and time to a PC computer or log printer.
  • Digital microprocessor control automatically monitors and stores faults from the last four weeks.
  • USP-compliant conductivity measurement with temperature compensation can be switched on or off.
Warranty: 1 year, parts and labor
Specifications

Specifications

0.055µS/cm
3/4 in.
Up to 2 L/min.
1 to 5 ppb
<1
2°C to 35°C
372 mm
615 mm
330 mm
90/240 V
<0.1 CFU/mL
Integrated
18.2 Mohm.cm at 25°C
0.1 to 6 bar
0.1 kW
14.6 in.
24 in.
13 in.
90/240 V 50/60 Hz
50/60 Hz
Videos
SDS
Documents

Documents

Product Certifications